KRQ-250A Growth Cabinet
Applicable to fundamental research and quality inspection in agricultural and forestry research, growth tissue culture of plants, life science and pharmacy, etc.
KRQ-250A Growth Cabinet
Purposes
Applicable to fundamental research and quality inspection in agricultural and forestry research, growth tissue culture of plants, life science and pharmacy, etc.
Features:
1.CPU microprocessor is adopted for the liquid crystal controller of temperature and humidity for PID control, with precision and reliable temperature and humidity data.
2.Ambient temperature monitoring device in the temperature controlling system can automatically control working condition of the cooling system, to effectively ensure long-term continuous operation of the equipment.
3.Pressure balancing control is utilized for the cooling system, which not only improves performance indicators, but also reduces power consumption.
4.Main parts are all high-quality imported ones, to ensure stability and reliability of the equipment.
5.The power off memory function ensures continuity of experiment.
6.Reasonable water treatment system truly realizes the drainless state.
7.Over-temperature alarm and temperature limiting protection function, etc.
Specification:
Model | KRQ Series of Growth Cabinet | ||||||
KRQ-250A | KRQ-250B | KRQ-250P | KRQ-300 | KRQ-300P | KRQ-400 | KRQ-400P | |
Volume | 250L | 300L | 400L | ||||
Power voltage | 220V 50HZ | ||||||
Temperature controlling range | No light:0℃~50℃ Under light:10℃~50℃ | ||||||
Temperature/Humidity resolution | 0.1℃/1%RH | ||||||
Temperature fluctuation | ±0.5℃ | ||||||
Temperature uniformity | ±2.0℃ | ||||||
Illuminance | 5 grades(adjustable) | 6 grades(adjustable) | |||||
0-12000LX | 0-15000LX | 0-20000LX | 0-24000LX | ||||
Program control(hour x segments x period) | N/A | 99x30x99 | N/A | 99x30x99 | N/A | 99x30x99 | |
Inner dimension(mm) | 500x600x850 | 480x580x1080 | 520x670x1150 | ||||
Outline dimension(mm) | 820x820x1565 | 800x810x1750 | 840x900x1850 | ||||
Carrier(standard configuration) | 3 pieces | 4 pieces |
Note: “A” refers to two sides under light; and other models have three side under light.
Conditions of no-load performance test: ambient temperature: 20℃, and ambient humidity: 50% RH;
Operating conditions for equipment: ambient temperature: 5-30℃。
Options:
1.Printer and USB interface.