KDN Semi-automatic Kjeldahl Nitrogen Analyzer
The instrument is composed of digestion furnace (different types for selecting) and azotometer distillatory.The digestion furnace (numeric display model) is controlled by temperature control meter. It adopts infrared radiation heating. Insert the
Product Name: Semi-automatic Kjeldahl Nitrogen Analyzer
Model: KDN
Principle
Measure principle steps of the instrument are shown as follows (Kjeldahl nitrogen determination principle)
sampling-digestion-distillation-titration-calculation
The instrument is composed of digestion furnace (different types for selecting) and azotometer distillatory.
The digestion furnace (numeric display model) is controlled by temperature control meter. It adopts infrared radiation heating. Insert the sample into the heating furnace to get the best heating effect.
Through the drainpipe and the air-bleed three-way, poisonous gases (e.g. SO2) emitted from the digestion tube could be taken to the cloaca by water (the gases could also be recycled by vacuum pump and recycle bottle) which could control the emitting of poisonous gases and save the installation of ventilated case. Samples could be digested thoroughly in about 40-90 minutes (according to the amount of nitrogen).
Main technical parameters
Measure variety | foodstuff, feedstuff, foods, dairy products, drink, soil, water, medicine, precipitate, chemical |
Working mode | semi-automatic |
Measure quantity | digest 4~20 pieces of tubes at the same time, adopt single distillation |
Measure range | 0.1~200mgN (nitrogen content 0.1%-99%) |
Recovery rate | ≥99% (relative error, include digestion process) |
Repetition rate | relative standard deviation< ± 1% |
digestion time | 40-90 minutes(According to different samples, time required is also different.)
|
distillation time | 5-15 minutes(According to different samples, time required is also different.)
|
Power | 4 holes---1000W, 8 holes---1500W, 20 holes---2000W Distillation: 1000W |
Attention: This instrument is a distillation unit; it should be used together with digestive furnace. Models of digestive furnace are: KDN-04C, KND-08C, KDN-20C, For more details, please contact us.